Abstract
The liquid chromatography–tandem mass spectrometry (LC-MS)/MS methodology was used to develop and validate a method for detecting erythromycin and olaparib in human plasma. Antibiotics such as erythromycin and olaparib fall into this category. Liquid chromatography is used to separate stationary and mobile phases based on differences in their affinities as well as to remove unwanted contaminants. It improves repeatability, sensitivity, resilience, and low-level protein detection. A C18 (C18, 5 m, 100×4.6 mm) column is utilized for high resolution and peak area. The calibration curve is created using linear regression. Internally, telmisartan is utilized as a benchmark. The flow rate of the mobile phase is 0.5 mL/min. Erythromycin and olaparib have mass-to-charge ratios of 735.43–115.97 and 435.08–102.04, respectively. Erythromycin in combination with olaparib resulted in a 98% recovery rate. The precision and accuracy of the results determined by interday QC samples are within acceptable limits. There was no evidence of instability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Asian Journal of Pharmaceutical and Clinical Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.