Abstract

A simple and selective LC method is described for the determination of Aspirin and Clopidogrel in tablet dosage forms. Chromatographic separation was achieved on a c18 column along with mobile phase consisting of a combination of fifty five volumes of Mixed Phosphate Buffer and forty five volumes of Acetonitrile with detection of 235 nm. Linearity was observed in the range 20-60 μg/ml for Aspirin (r2=0.998) and 10-30 μg /ml for Clopidogrel (r2 =0.998) for the amount of drugs estimated by the projected ways was in smart agreement with the label claim. The proposed methods were validated. The accuracy of the methods was assessed by recovery studies at three different levels. Recovery experiments indicated the absence of interference from commonly encountered pharmaceutical additives. The method was found to be precise as indicated by the repeatability analysis, showing %RSD trials. All statistical data proves validity of the ways and may be used for routine analysis of pharmaceutical dosage form.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.