Abstract

Objective: The aim of this study is to develop a simple, precise and accurate analytical method of deoxyarbutin in anhydrous emulsion system preparation.
 Methods: The analysis was conducted using high-performance liquid chromatography (HPLC). Chromatographic analysis was carried out using a reversed phase-C18 column. The mobile consists of two phases methanol and water (60: 40 v/v) at a flow rate of 1.0 ml/min. The determinations were performed using UV detector set at 225 nm. All validation procedures were added with hydroquinone as an internal standard.
 Results: The method showed coefficient correlation is 0.9978, relative standard deviation (RSD) smaller than 2%, Limit of Detection (LOD) and Limit of Quantitation (LOQ) are 0.599 µg/ml and 1.817 µg/ml respectively. The total amount deoxyarbutin in anhydrous emulsion preparation is 1.964+0.02 % with 98% recovery percentage.
 Conclusion: The developed HPLC analytical method meets the validation criteria made by International Conference on Harmonisation (ICH).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.