Abstract
Quantification of uranium in human urine is a valuable technique for assessing occupational and public exposure to uranium. A reliable method has been developed and validated in the ARPANSA Radiochemistry Laboratory by means of standard radiochemical separation and purification techniques and measurement using high-resolution alpha spectrometry. This method can be used to evaluate the levels of naturally occurring 234U, 235U and 238U in urine. Method design and validation is the process of defining an analytical requirement, and then confirming that the method under consideration has performance capabilities consistent with what the application requires. The method was designed to measure levels down to 2 mBq/day of total uranium, corresponding to approximately 1/100th of the annual committed effective dose of 20 mSv. Validation tests were developed to assess selectivity, accuracy, recovery and quantification of uncertainty. The radiochemical recovery of this method was measured using 232U tracer. The typical minimum detectable concentration for total uranium for 24-h urine samples is approximately 0.6 mBq/day or 0.019 μg/day.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.