Abstract

The paper proposes a method and its realization for stabilizing the characteristics of a multichannel position-sensitive spectrometer. The device uses a new design of the semiconductor strip detector, which allows a simultaneous injection of stable in time electric charges on all strips. This solution enables the electronics to track changes in the conversion characteristics of all electronic channels including detectors strips, and correct them during the spectrometer operation. In addition, the proposed detector design makes it possible to control the signal propagation along the strips, thus providing an on-line check of the integrity of strips metallization and the quality of their connection with the readout electronics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.