Abstract
Background:Tumescent local anaesthesia with prilocain can lead to clinically significant methemoglobin levels. New generation multiple wavelength pulse oximeters (e. g. Masimo Radical 7®) can measure methemoglobin levels. Methods:In this prospective observational study we compared the venous methemoglobin levels and the corresponding pulse oximetric values of the Radical 7® in patients undergoing tumescent local anaesthesia for liposuction procedures. The measurements were performed in Hanseklinik, Luebeck, Germany between 2008 and 2011.Results:In 133 patients, we measured a maximum methemoglobin level of 18 per cent. In a Bland-Altman analysis we found a mean bias of +2.2 % (-4.1 to 8.4 limits of agreement) for pulse oximetric values compared to hemoximetry. Conclusion:Pulse oximetric measurement of methemoglobin is an early-warning tool for the detection of clinically significant methaemoglobinemia in patients with tumescent local anaesthesia.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.