Abstract
Three American products and one Canadian product were examined for content uniformity and particle size distribution. The results showed that not all products performed equally well. Some of the products exhibited high sprays early in the canister lifetime and all products demonstrated loss of prime. The particle size distributions were determined using the Andersen cascade impactor (USP Induction Port) and the fine particle fraction was determined using the twin impinger. The results showed that three of the four products had similar particle size distribution profiles. Both the Andersen cascade impactor and the twin impinger yielded the same trends in the amount of drug substance delivered to the fine particle fraction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.