Abstract

Compact and minimally invasive scanning fiber endoscopy probes with micron-level resolution have great potential in detailed tissue interrogation and early disease diagnosis, which are key applications of confocal reflectance imaging at visible wavelengths. State-of-the-art imaging probes commonly employ refractive lens triplets or gradient refractive index (GRIN) lenses as the micro-objective. However, off-axis aberration emerges as a critical factor affecting resolution, especially at the extremities of the imaging field. In response to this challenge, we propose what we believe to be a novel design integrating a metasurface with the GRIN micro-objective to address optical aberrations during beam scan. The metasurface acts as a corrector element for optical aberrations in a fiber-scanning endoscope using the same fiber for excitation and collection. Modeling such hybrid refractive-metasurface designs requires the coupling of simulation techniques across macroscale and nanoscale optics, for which we used an Ansys simulation workflow platform. Operating at a wavelength of 644 nm, this metaoptical element serves as a thin and compact aberration correction surface, ensuring uniform resolution across the entire imaging field. Experimental results from our scanning fiber endoscopy system demonstrate a notable enhancement in optical performance both on-axis and off-axis, achieving a resolution of 3 µm at the center of the imaging field. Impressively, the resolution experiences only a modest degradation by a factor of 0.13 at the edge of the field of view compared to the center.