Abstract

High-Tc superconducting quantum interference device (SQUID) system for detection of magnetic foreign matter in industrial products was developed. There is a possibility that ultra-small metallic foreign matter has been accidentally mixed with industrial products, such as lithium ion batteries. The outer dimension of metallic particles less than 100 microns cannot be detected by conventional X-ray imaging. Therefore, we developed a detection system based on high-Tc SQUID microscopes with a high-performance magnetic shield. Using SQUID microscopes with a 0.5 mm-thick vacuum window was proposed. This design enables the SQUID to approach an object to be measured as close as 1 mm and enhances the sensitivity. A new magnetic shield with sleeves was carefully designed and built. As a result, we could successfully measure a small iron particle with 100 μm. This detection level was hard to achieve by conventional X-ray detection methods.