Abstract

This study aimed to evaluate the effectiveness of spectral computed tomography (CT) mono-energy imaging combined with metal artifact reduction software (MARs) for metal implant artifact reduction using a phantom. A quantitative standard phantom with 9 cylinders was used to simulate the attenuation of the different tissues of the human body around the metal implant. Groups A and B were divided according to conventional CT scan mode and spectral CT scan mode. Three sets of reconstructed images, including 120 kVp-like + MARs images, mono-energy images (MonoE), and MonoE + MARs images, were generated after spectral CT scanning. High-attenuation artifacts and low-attenuation artifacts were observed around the coil in the images of groups A and B. The CT values (Hounsfield unit) and standard deviation (SD) values of the artifacts were measured, and the artifact index and hardening artifact removal rate were calculated. Compared to conventional poly-energy CT images, for high-attenuation and low-attenuation artifacts, the artifact indices of 120 kVp-like + MARs, MonoE, and MonoE + MARs images were all reduced significantly. The hardening artifact removal rates of the high-attenuation and low-attenuation artifacts of 120 kVp-like + MARs images were 82% and 92%, respectively. The hardening artifact removal rate of the high-attenuation and low-attenuation artifacts of MonoE and MonoE + MARs images increased with the mono-energy level. Spectral CT using the 120 kVp-like + MARs, 110-140 keV MonoE, and MonoE + MARs reconstruction methods can reduce metal implant artifacts in varying degrees. MonoE + MARs reconstruction was the best method for reducing metal artifacts.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call