Abstract

The objective of this study was to analyze postoperative outcomes in a group of patients who underwent metacarpophalangeal (MCP) arthroplasty using a pyrocarbon prosthesis for noninflammatory arthritis. An analysis of 44 MCP joint arthroplasties in 30 patients with >2 years of follow-up over a 12-year period was reviewed. The mean age was 63 years. The primary operative indication was pain and stiffness from osteoarthritis refractory to nonsurgical management. At a mean follow-up of 6 ± 3 years, 8 (18%) joints underwent reoperation, including 5 (11%) that underwent revision arthroplasty. The 2- and 5-year rates for survival free of revision arthroplasty were 95% and 93%, respectively. One (2%) operation was complicated by intraoperative fracture. Postoperative complications occurred in 8 (18%) fingers and included ligament/tendon rupture (n= 3) and instability (n= 2). There was significant postoperative improvement in pain levels, MCP arc of motion, pinch strength, and grip strength. At a mean 5 years of radiographic follow-up, 7% had progressive implant instability because of grade 3 or greater loosening. No joints experienced implant instability from progressive subsidence. Metacarpophalangeal arthroplasty using a pyrocarbon implant for osteoarthritis demonstrates an 7% revision rate at 5 years after surgery. Complications lead to reoperation in 1 of 5 arthroplasties. Radiographic evidence of implant instability was uncommon. Overall, patients experienced predictable pain relief and improvements in their range of motion and pinch strength. Therapeutic IV.