Abstract

Two novel potentiometric sensors that are highly selective to Hg2+ ions are described. These are based on the use of 5,5'-dithio-bis(2-nitrobenzoic acid) (DTNB) and tricyclazole (TCZ) as neutral carriers in plasticized poly(vinyl chloride) membranes. Fast Nernstian responses are obtained for Hg2+ ions over the concentration ranges 7.0 x 10(-6) - 1.0 x 10(-2) and 7.7 x 10(-6) - 1.0 x 10(-2) mol l(-1) at pH 1.8 - 3.3 with lower detection limits of 5.0 x 10(-6) and 5.6 x 10(-6) mol l(-1) (approximately 1 microh ml(-1)) and calibration slopes of 30.0 and 29.7 mV decade(-1) with DTNB- and TCZ-based membrane sensors, respectively. Validation of the assay method reveals good performance characteristics, including long life span, good selectivity for Hg2+ ions over a wide variety of other metal ions, long term response stability, and high reproducibility. Applications for direct determination of mercury in hazardous wastes including dental amalgam, mercury bulbs, and fluorescent lamps give results with good correlation with data obtained using cold vapor atomic absorption spectrometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.