Abstract

Pollution due to heavy metal ions, including mercury, has become a major issue because of their toxicities. It is required to monitor mercury levels in aqueous media using fast and selective methods with high accuracy. Ellipsometry is a promising technique for instance when it's combined with the plasmon resonance phenomena. We reported a biosensor system available for qualitative/quantitative determination of mercuric ions in aqueous media where both the spectrophotometric ellipsometry and oligonucleotide recognition elements were used. A single step assay using both a linear (ProbeL) and a hair-pin (ProbeH) type oligonucleotide probe as a recognition element, in addition to a sandwich-type (ProbeLS) assay were developed and compared. The detection limits were 0.23 nM, 0.03 nM and 0.15 pM for ProbeL, ProbeH and ProbeLS, respectively. The detection range was between 0.05nM and 100nM Hg2+ for all assays proposed herein.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.