Abstract

A single flow channel analysis system that employs the formaldehyde-S(IV)-pararosaniline color reaction has been developed for the determination of aqueous S(IV). The system employs a passive cation-exchange membrane reactor for the introduction of base and a pressurized porous membrane reactor for the introduction of acidic pararosaniline. The detection limit is 1.6 x 10/sup -7/ M S(IV); the sample size is 200 ..mu..L, and the throughput rate is 15/h. Results for cloudwater samples are presented. 15 references, 3 figures, 1 table.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.