A semi‐automated method to determine pK a values of drugs and chemicals spectrophotometrically is described. The method uses the capabilities of a HPLC with a diode array detector (DAD) as a flow injection apparatus and a binary solvent delivery system. The advantages of this method are low sample consumption (less than 0.1 mg compound needed), high sample throughput, easy set up, high sensitivity, and low cost, precision within a 0.2 unit. Experimental pK a values obtained for over thirty model compounds are consistent with literature values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.