Abstract

Proton therapy is a technique used to deliver a highly accurate and effective dose for the treatment of a variety of tumor diseases. The possibility to have an instrument able to give online information could reduce the time necessary to characterize the proton beam. To this aim we propose a detection system for online proton beam characterization based on the Medipix2 chip. Medipix2 is a detection system based on a single event counter read-out chip, bump-bonded to silicon pixel detector. The read-out chip is a matrix of 256×256 cells, 55×55 μm 2 each. To demonstrate the capabilities of Medipix2 as a proton detector, we have used a 62 MeV flux proton beam at the CATANA beam line of the LNS-INFN laboratory. The measurements performed confirmed the good imaging performances of the Medipix2 system also for the characterization of proton beams.