Abstract
Vocal cord paralysis is a disease that can cause voice disability and aspiration problems. Expanded polytetrafluoroethylene (ePTFE; Gore-Tex) has been widely used in cardiovascular and plastic surgery; however, its biocompatibility and safety have not been established precisely in the larynx. In this study, the biocompatibility and safety of ePTFE used in vocal cord medialization in the rabbit larynx were assessed. Type 1 thyroplasty with Gore-Tex was performed on one side of the larynx in eight rabbits. A sham procedure was applied on the contralateral side of their larynx. The rabbits were sacrificed after 6 months, and macroscopic and microscopic evaluations and comparisons were performed. No significant inflammatory response to Gore-Tex or displacement of the implant was seen. There was only a tiny fibrotic capsule surrounding the material in each specimen. Intense eosinophilic cell infiltration was seen surrounding one rabbit specimen. Gore-Tex seems to be a biocompatible, stable, and reversible material that can be safely used in laryngeal framework surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.