Abstract

As most risk factors for anastomotic complications (AC) in rectal cancer patients appear to be noncorrectable, it is needed to find the correctable causes. Additionally, the outcomes of indocyanine-green fluorescence imaging (IFI) and robot-stapled anastomosis have yet been undetermined. This study retrospectively analyzed 968 consecutive patients with rectal cancer, who underwent curative robot-assisted anterior resections between 2010 and 2018. IFI parameters and stapling features in the surgical records were reviewed, and reconfirmed. AC occurred in 54 patients (5.6%), 34 (3.5%) with anastomotic leakage (AL) and 24 (2.5%) with anastomotic stenosis (AS). Mechanotechnical faults including defective stapling configurations, including angles lesser than or equal to 150° and outer deviation (more than half from the center of the circle) of linear staples, between the two linear staples were independently associated with AL (P < .001 each). IFI significantly reduced AL rate (2.5% vs 5.3%, P = .029) and AS rate (2% vs 18.8%, P = .006), respectively. Robot linear stapling enabled to maintain the obtuse angle during consecutive staplings and reduced console time. AL and AS were independent risk factors for disease-free survival (P = .02) and local recurrence (P = .03), respectively. AC were associated with some correctable causes, namely, mechanotechnical errors and lack of use of IFI.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call