Abstract
The measurement of the ultrasonic output power emission from transducers is a very important subject because strong ultrasonic power may be harmful to human health during medical diagnosis and therapy. Conventional techniques for measuring ultrasonic power such as radiation force balance method and calorimetry method provides no information on power distribution, laser interferometry method is very expensive and time-consuming. We have demonstrated in previous research that the mechanoluminescent (ML) film can be used as a sensor for visualizing the ultrasonic power distribution of a transducer with a frequency of 20 MHz. In this research, the ultrasonic power distribution of a transducer with a frequency of 6 MHz was also visualized. The results suggest that this method can be used for visualizing the ultrasonic power distribution of transducers with a wide range of frequency.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IOP Conference Series: Materials Science and Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.