Abstract

ABO and Rh (D) groups of 6403 blood samples were assessed on an 8/9-channel autoanalyser in the Serology Department of the London Hospital; the results were independently checked at the Regional Blood Transfusion Centre, Brentwood, using the routine methods for grouping donor blood. Results of this comparative study are given and instances are described in which anomalous results or incorrect groupings occurred; the possible causes are discussed. The 8/9-channel automated blood group analyser is evaluated in terms of routine hospital laboratory practice.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.