Abstract

An automated assay of plasma prekallikrein is described. Prekallikrein was converted to kallikrein with Pseudomonas aeruginosa elastase, and the hydrolytic activity of kallikrein to H-D-Pro-Phe-Arg-paranitroanilide subsequently measured. The conversion was complete within 8 minutes and the amidolytic activity remained stable at least another 10 min at 37 ° C. This method worked in plasma deficient in Hageman factor (blood coagulation factor XII). Using anti-prekallikrein antibody and plasma deficient in prekallikrein, the amidolytic activity generated in normal plasma was identified as due to kallikrein. With plasma samples, the coefficients of variation (CV) for multiple measurements within run ( n = 10) and between run ( n = 10) were as low as 5.0% and 6.6%, respectively, and the minimum measurable concentration of prekallikrein in plasma was 10% of the normal level.