Abstract

Tests carried out on the Hoffmann- Vidal external fixator by means of holographic interferometry are reported. These tests refer exclusively to the quality of functioning of the joints connecting transfixing pins and side bars. It is a question of 1) evaluating the behavior of the ball joint subjected to realistic loads, 2) verifying the rigidity of a lightened joint bowl as compared to a heavier one, and 3) comparing the respective resistance of two similar joint models to a rotation of the bar around its axis. The main findings are the remarkable elastic behavior of the ball joint, the effectiveness of the lightened bowl, and the fact that two cousin models may behave quite differently with respect to their resistance to axial slipping rotation of the bar. The interferometric technique is thus expected to have further applications in similar evaluations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.