Abstract

In joint replacement, cyclic motion at the bone-prosthesis interface is considered a precursor to component loosening. This study characterized the mechanical stability of 13 total knee arthroplasties harvested postmortem after an average time in situ of 10.3 years. With loads applied to the medial and then the lateral tibial plateau, motion between the tibial component and underlying bone was measured with extensometers. The amount of motion between the tibial component and underlying bone under medial and lateral loads of 500 N and then twice body weight was typically less than 20 μm. Tray depression under load application and the liftoff on the contralateral side indicated that the tibial stems limited implant rotation and that implant fixation did not deteriorate with time in situ.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.