Abstract

Although new generation alumina ceramics have exhibited a reduced incidence of fracture, concern still persists about the behavior of ceramic acetabular liners under impact conditions. The objective of this study was to explore whether fracture of a new generation alumina ceramic liner was likely to occur in vivo. Ceramic liners were impacted with forces of 23, 21, 15, and 12 kN (n = 3 at each force). At 23 kN, all 3 ceramic liners fractured on the first impact; at 12 kN none of the ceramic liners fractured after 20 impacts. The threshold force of 12 kN is large in comparison with estimated physiologic forces on the hip during falls or stumbling, suggesting that ceramic liner fracture is not a definite consequence of liner impact.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.