Abstract

This paper deals with the mechanical behavior of polyether–ether–ketone (PEEK) under impact loading. PEEK polymers are the great interested in the field of medical implants due to their biocompatibility, weight reduction, radiology advantage and 3D printing properties. Implant applications can involve impact loading during useful life and medical installation, such as hip systems, bone anchors and cranial prostheses. In this work, the mechanical impact behavior of PEEK is compared with Ti6Al4V titanium alloy commonly used for medical applications. In order to calculate the kinetic energy absorption in the impact process, perforation tests have been conducted on plates of both materials using steel spheres of 1.3g mass as rigid penetrators. The perforation test covered impact kinetic energies from 21J to 131J, the equivalent range observed in a fall, an accidental impact or a bike accident. At all impact energies, the ductile process of PEEK plates was noted and no evidence of brittle failure was observed. Numerical modeling that includes rate dependent material is presented and validated with experimental data.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.