Abstract

We have evaluated the resistance to displacement of six stable methods of fixation of a sagittal split ramus osteotomy (SSRO) in the mandibular advancement with counterclockwise rotation. We tested 60 synthetic hemimandibles in six groups of 10 each: Group I – fixation with a straight four-hole 2.0mm miniplate; Group II – a straight six-hole 2.0mm miniplate; Group III – two straight 2.0mm four-hole miniplates; Group IV – an eight-hole 2.0mm (grid plate); Group V – a 2.0mm four-hole straight miniplate and 2.0×12mm bicortical screw; and Group VI – a straight four-hole 2.0mm locking miniplate. We applied a linear force in the region between the canine and the first premolar using a universal testing machine (EMIC- DL2000) with a loading cell of 10 KN. The loads at 1, 3, and 5mm displacement were recorded (N) and the data transmitted from the load cell to a computer. Results were analysed using analysis of variance (ANOVA) (p<0.001) and the Tukey post-test for comparison of the significance of the differences between the groups. For the three degrees of displacement, fixation with two straight 2.0mm plates and with the grid plate gave higher load values.