Abstract

To compare the standard DFNS devices for assessment of mechanosensory function to a low resource tool, the Sorri-Bauru-monofilaments. Mechanical detection thresholds (MDT), pain thresholds (MPT), and suprathreshold pinprick ratings (pain sensitivity: MPS) were measured over cheek, hand dorsum, and fingertip in 13 healthy subjects (7 female, aged 21-44 years). Mechanical detection threshold was assessed with DFNS standard glass monofilaments (0.25-512 mN, 0.5 mm tip) and nylon monofilaments (Sorri-Bauru; 0.5-3000 mN). MPT was assessed with DFNS standard cylindrical probes (8-512 mN, 0.25 mm tip), Sorri-Bauru monofilaments, and with ramped stimuli using an electronic von Frey aesthesiometer (10 mN/s or 100 mN/s, 0.20 mm tip). MPS was measured in response to stepped and ramped pinpricks (128 and 256 mN). Mechanical detection thresholds were the same for DFNS and Sorri-Bauru monofilaments. For MPT, Sorri-Bauru filaments yielded lower values than PinPricks over face but not hand. Pain thresholds were higher at all test sites for ramped than stepped pinpricks (P < 0.01). Suprathreshold ratings were lower for ramped than stepped pinpricks (P < 0.05). Sorri-Bauru filaments are acceptable substitutes for DFNS standards in estimating tactile sensitivity, but are not consistent with standard probes for pinprick sensitivity because of their nonstandardized tips. Ramped stimuli overestimated MPT and underestimated MPS due to reaction time artefacts and therefore need their own normative values.