Abstract

Using a mechanical testing procedure, various fixator constructs were tested in vitro. In addition, the influence of the passive soft tissue structures on the fixation stiffness was determined. An increased number of Schanz' screws or Kirschner wires led to a comparable increase in stiffness than that observed with an increasing screw or wire diameter. In consequence, larger diameters should be preferred over an additional screw or wire where clinically applicable. With diaphyseal telescoping rods only the axial stiffness decreased. As expected, large ring diameters as well as titanium wires reduced stiffness components. Bracing the outer rings caused a reduction of the overall stiffness. Asymmetric pre-tensioning of the K-wires resulted in a significant reduction of tension in the neighboring wire. Removal of the soft tissues reduced stiffness to a similar extend as experienced in a fibula defect situation. The study demonstrates the correlation between design parameters, passive soft tissues and fixation stiffness and presents guidelines for an optimized fixator design.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.