Abstract

The most common means of sternal closure after sternotomy is stainless steel wire cerclage. These wires, while inexpensive and simple in design, are known to be associated with low strength and sternal dehiscence. In this biomechanical analysis, we compare single sternal wires, double sternal wires, and a novel sternal closure device we have designed to mitigate sternal dehiscence. The device uses polymer grommets at the sternal interfaces to distribute load over a large surface area of bone. Samples of each closure device were installed in a bone model and distracted at a rate of 10 mm/min while tensile forces were continuously measured and compared. Single wires generated the lowest stiffness and strength values, followed by the double wires. The novel device demonstrated significantly higher stiffness and strength at all displacements compared with the single and double wires. Clinical use of this device may result in meaningful reduction in complications associated with the use of standard sternal wires such as sternal separation and fracture.