Abstract

Next-generation X-ray observatories, such as the Lynx X-ray Observatory Mission Concept or other similar concepts in the coming decade, will require detectors with high quantum efficiency (QE) across the soft X-ray band to observe the faint objects that drive their mission science objectives. Hybrid CMOS Detectors (HCDs), a form of active-pixel sensor, are promising candidates for use on these missions because of their fast read-out, low power consumption, and intrinsic radiation hardness. In this work, we present QE measurements of a Teledyne H2RG HCD, performed using a gas-flow proportional counter as a reference detector. We find that this detector achieves high QE across the soft X-ray band, with an effective QE of $94.6 \pm 1.1 \%$ at the Mn K$\alpha$/K$\beta$ energies (5.90/6.49 keV), $98.3 \pm 1.9 \%$ at the Al K$\alpha$ energy (1.49 keV), $85.6 \pm 2.8 \%$ at the O K$\alpha$ energy (0.52 keV), and $61.3 \pm 1.1 \%$ at the C K$\alpha$ energy (0.28 keV). These values are in good agreement with our model, based on the absorption of detector layers. We find similar results in a more restrictive analysis considering only high-quality events, with only somewhat reduced QE at lower energies.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call