Abstract

The application of the magnitude-squared coherence (MSC) spectrum as a measure of the degree of organization of the cardiac electrical activity is explored. The MSC spectrum is a frequency-domain measure of the linear relationship between two signals. In the work described the two signals are two bipolar electrograms from either acutely placed catheter(s) or automatic implantable cardioverter/defibrillator electrodes. It is shown that the MSC is a dimensionless (no units), real-valued spectrum that is always in the range of zero to unity. The case of zero is found at frequencies where there is no linear relationship between the signals, and the case of unity implies a linear, noise-free relationship. The way the MSC spectrum is normalized makes it insensitive to gain or gain differences between the two signals. Example MSC spectra are presented and discussed. Striking differences in the spectra for fibrillatory and nonfibrillatory rhythms are seen.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.