Abstract

Single microphone noise reduction (NR) in hearing aids can provide a subjective benefit even when there is no objective improvement in speech intelligibility. A possible explanation lies in a reduction of listening effort. Previously, we showed that response times (a proxy for listening effort) to an auditory-only dual-task were reduced by NR in normal-hearing (NH) listeners. In this study, we investigate if the results from NH listeners extend to the hearing-impaired (HI), the target group for hearing aids. In addition, we assess the relevance of the outcome measure for studying and understanding listening effort. Twelve HI subjects were asked to sum two digits of a digit triplet in noise. We measured response times to this task, as well as subjective listening effort and speech intelligibility. Stimuli were presented at three signal-to-noise ratios (SNR; –5, 0, +5 dB) and in quiet. Stimuli were processed with ideal or nonideal NR, or unprocessed. The effect of NR on response times in HI listeners was significant only in conditions where speech intelligibility was also affected (–5 dB SNR). This is in contrast to the previous results with NH listeners. There was a significant effect of SNR on response times for HI listeners. The response time measure was reasonably correlated (R142 = 0.54) to subjective listening effort and showed a sufficient test–retest reliability. This study thus presents an objective, valid, and reliable measure for evaluating an aspect of listening effort of HI listeners.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call