Abstract

Background Vascular compression is important for deep vein thrombosis screening. However, pressure analysis of ultrasound vessel models has not been performed. Therefore, we compared the human popliteal vein and several ultrasound vessel models at 50% compression. Methodology Four major ultrasound vascular models used in Japan and the popliteal vein of one subject constituted our measurement targets. Using a pressure-sensitive measuring device, the compressive force required to shorten the vessel diameter by 50% was determined. Results The compression force that shortened the popliteal vein by 50% was measured to be 191 ± 65 g. The blue phantom, ultrasound CV Pad II, ultrasound training block, and UGP-GEL required compression force of 701 ± 8 g, 265 ± 12 g, 697 ± 20 g, and 745 ± 15 g, respectively. The compression force for the ultrasound training block was 2.6 times higher than that for the ultrasound CV Pad II. The gel material around the vessels was the same; however, different vascular tubes required 2.6 times higher compression force. Conclusions This study showed that the objective numerical values of the compressive force were required to compress an ultrasound vascular model. Reproduction of the compressibility of veins required either removing the vascular structure or using thin tubing material.