Abstract
Movements of soft tissues surrounding the oral cavity, especially lips and cheeks, have a strong influence on mastication and phonetics. They also influence the relationship between a denture and its oral environment. The purpose of this study was to develop a three-dimensional measuring system for soft tissue movement. This system consisted of two video trackers placed stereographically and a computer. In addition, one video tracker was connected for measuring mandibular movements. The accuracy of this system was evaluated using computerized XYZ pulse stage. The resolution of this system was 0·10×0·10×0·10 mm, when the measurement was carried out in the area of the 24·0 (X)×20·0 (Y)×20·0 (Z) mm with a working distance of 500 mm and a frequency of 120 Hz. In the present study, the lip movements of a dentulous subject with mandibular movements during chewing peanuts were analysed using this system. The new system demonstrated its value for analysing soft tissue movement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.