Abstract

The aim of our developments is three-dimensional in vivo recording of those orthodontic force systems inducing tooth movements during treatment with fixed appliances. The concept presented here is the first to permit the forces and torques of these statically multiply undetermined systems to be recorded in vivo. For this purpose the force systems transmitted to the teeth from the archwire are isolated from the respective tooth by means of divisible special-design brackets and introduced into a 3D force–torque sensor via a gripping appliance. The sensor is fixed with a purpose-developed device relative to the patient’s dental arch. The patient’s head is positioned relative to the system by means of a bite fork as well as a forehead and chin support. Electrical measurement of the mechanical quantities is carried out by a six-axis force–torque sensor with semiconductor strain gauge elements, an electronical evaluator and a mobile measuring computer (PC). Extensive calibration of the sensor system has shown that the measuring uncertainty of the electrical measuring is less than 2%. Precise spatial fixing of bracket slot and archwire in the therapeutic position is crucial to the measuring accuracy of the system, as even minimum displacements affect the force system to be measured. Movements of the measuring system up to 0.04 mm result from a therapeutic force of 1.5 N. The results of extensive in vitro studies have already demonstrated that the system developed by us is suitable for the specified in vivo measuring function.