Abstract

Intra-abdominal pressure measurement is essential to the diagnosis of patients with intraabdominal hypertension or abdominal compartment syndrome. The most common method for measuring intra-abdominal pressure (IAP) is the intravesicular or "bladder" technique, which requires electronic monitoring technology not available on the typical surgical ward. Herein we describe and validate a simple, rapid screening method for bedside IAP measurement using the patient's indwelling urinary catheter and a readily available intravenous tubing extension. Validation of this technique across the clinically important IAP diagnostic spectrum demonstrated acceptable bias (1.6 mm Hg; 95% confidence interval 1.4-1.8) with limits of agreement of 0.36 to 2.8. This demonstrates good agreement between the two IAP methods and validates the bedside technique as a simple, cost-effective, and reproducible method for screening IAP measurements outside of the critical care setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.