Abstract
A performance study was performed for CdZnTe coplanar grid (CPG) detectors when used as γ-ray spectrometers. The detectors have the crystal volumes of 1, 1.6875 and 2.25cm3, respectively. Time stability of each CdZnTe CPG detector was investigated in a long-term operation (time span of 0.25 to about 100h). The spectroscopic performances were measured at different bias voltages and at various photon energies ranging from 59.6keV (241Am) to 1332.5keV (60Co) for each detector, and evaluated by using the following parameters: energy resolution in FWHM, peak tailing in peak-to-valley (P/V) ratio and in FWHM/FW.25M ratio, and photofraction using the acquired γ-ray spectra. No polarization effect was observed in terms of count rate, energy resolution and peak centroid shift. The obtained results indicate that better time stability and excellent spectroscopic performances of the present CdZnTe CPG detectors are shown for a room temperature γ-ray spectroscopy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.