Abstract

Characteristics with respect to light scattering intensity (LSI) were measured on latex spheres and on sphered and unsphered red blood cells (RBC). From the discrepancies in LSI signals from polymer spheres and RBCs it is obvious that latices cannot be used for absolute calibration of RBC sizes. Sphering of RBCs did not give rise to a closer resemblance. Monodisperse latex spheres produce extremely narrow size distribution histograms. Regarding these sharp peaks and the long-term stability, polymer spheres offer definite advantages if compared with blood controls to which preservatives were added. A shift from the RBC size distribution histogram, caused by gradually developing instrument drift, can be easily detected and corrected at an early stage.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.