Abstract

This study investigates the use of a dual-beam, optical null, FT-IR spectrometer to measure trace organic components in aqueous solutions in the combination band region 5000–4000 cm−1. The spectrometer may be used for both single- and dual-beam measurements, thereby facilitating comparison of these two modes of operation. The concentrations of aqueous solutions of urea and glucose in the ranges 0–40 mg/dL and 0–250 mg/dL, respectively, were determined by principal component regression using both modes. The dual-beam technique eliminated instrumental variations present in the single-beam measurements that must be taken into account when quantifying trace components from single-beam spectra. The data obtained with the dual-beam technique resulted in more stable calibration models based on principal component regression. These calibration models need fewer factors and yield lower prediction errors than those based on traditional single-beam data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.