Abstract
The measurement of beta/gamma and neutron dose components in mixed radiation fields is described involving combined OSL and TL measurements. The accuracy of the technique is investigated by irradiation of LiF:Mg,Ti samples with 14.1 MeV neutrons and a comparison of the OSL/TL results with the neutron fluence determined by Al activation. The precision of the measurements is 6.1 to 6.7 % (1 SD) for the OSL/TL technique compared to 7.3 % for the Al foils. The advantage of the technique is the simultaneous determination of the two dose components on the same samples of 7LiF:Mg,Ti in a single dedicated TL/OSL reader.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.