Surface dose was measured in an extrapolation chamber and was found to be less than previously reported. The slope of the build-up curve was less steep with the Clinac 4 than with 60-Co and much steeper with the Clinac 35 at 25 MV than with the betatron at the same energy. The gradient of the depth-dose curve in the build-up region is probably the best indicator of skin damage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.