Abstract

Beta-emitters 36Cl and 63Ni, and the electron capture radionuclide 55Fe were used to calibrate a secondary standardization instrument Quantulus-1220TM. An external standard technique was applied to assess the quenching level of samples. Nitromethane as an artificial quencher was used for sample preparation at different quenching levels. The dependence of the counting efficiency on the quenching level was determined. Samples of beta-emitters were re-measured with the triple-to-double coincidence ratio (TDCR) primary standardization instrument. The specific activities of beta-emitters were determined with the help of the software TDCRB-02 and compared with known values. It was found that for 36Cl and 63Ni the discrepancy was 0.9% and 2.7%, respectively, within the specified uncertainty limits. A new sample with 90Sr was prepared and measured with the TDCR device. In this case, the divergence from the expected value was 1.7%. Results suggest the suitability of the TDCR techniques for low-level counting.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.