Abstract

A new technique is presented for the measurement of neutron and/or gamma-ray coincidences. Separate neutron neutron, neutron gamma-ray, gamma-ray neutron, and gamma-ray gamma-ray coincidences are acquired with liquid scintillation detectors and a digital pulse shape discrimination (PSD) technique based on standard charge integration method. The measurement technique allows for the collection of fast coincidences in a time window of the order of a few tens of nanoseconds between the coincident particles. The PSD allows for the acquisition of the coincidences in all particle combinations. The measurements are compared to results obtained with the MCNP-PoliMi code, which simulates neutron and gamma-ray coincidences from from a source on an event-by-event basis. This comparison leads to good qualitative agreement.