Abstract

Measurements of central axis depth ionisation curves and ionisation at ionisation maximum were carried out in water, clear polystyrene and a commercially available 'solid water' phantom material. Flat and cylindrical chambers were used for electron beams of 5-10 MeV. Displacements for the cylindrical chambers were determined, indicating a recommended value of 0.55 times cavity radius (without perturbation corrections). The use of a single scaling parameter is considered for converting depth ionisation curves obtained in plastic to those in water. This is shown to be valid within 2 mm at these energies and for these materials. Ionisation ratios between water and polystyrene are presented, i.e. correction factors for converting ionisation readings in the plastic to readings in water required for electron dosimetry determinations. These show 3% differences on average at these energies, ionisation in water being higher. Variations were observed with chamber wall material and chamber type. Measurements in the solid water material showed it to be a better water substitute than polystyrene for electron dosimetry in this energy range, although still significantly different from a true water phantom.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.