Abstract
Two organic scintillators (EJ309 and stilbene) with their different detection chains and elaborate software were used to measure neutron spectra of AmBe and PuC sources. Method of shadow cone ensures minimization of scattered neutron influence. To separate neutrons and γ, a modern method of pulse shape discrimination was employed using fast digitizers. Calibration was performed by γ sources and three different methods for the identification of the Compton edge. Spectrometric systems also used different light output functions and unfolding methods. The same spectra were calculated using the MCNP6 code and the JENDL-5 library. The measured and calculated results were mutually compared, and also with the published results available in the literature. Generally, the results agree on the main peaks of neutron spectra but differ in their abundance. The lower limit for the ability to distinguish between neutrons and γ in spectrometry with organic scintillators was found at about 1.5 MeV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.