Abstract

As CT exams impart high doses to patients in comparison to other radiologist techniques, reliable dosimetry is required. In this work, dosimetry in CT beams was carried out in terms of air kerma index in air or in a phantom measured by a pencil ionization chamber, thermoluminescent (TL) detectors and radiochromic film. Calibration results showed the low energy dependence of all three dosimetric systems for the 100–120kV range, the very high uncertainty of the TL dosimeters in comparison to the other systems and high nonlinearity response in terms of air kerma of the radiochromic film. Measurements with the three systems in a 120kV CT protocol showed an acceptable agreement among the weighted air kerma index values, but TL dosimeters presented the highest uncertainties in the values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.