Abstract
Hydrogen chloride gas is measured by absorption in the second overtone band near 1.2microm using an InGaAsP diode laser. The strength and air-broadening coefficient of the H(35)Cl P(3) line are measured. The line strength is ~18% higher than the previous grating spectrometer measurement and suggests that the commonly used experimental value for the 3 ?0 HCl vibrational moment may be low. High-frequency two-tone FM detection is also used in this study to measure trace concentrations of HCl. Using a multimode laser, the sensitivity limit is 3-parts per million HCl in air at 50 Torr, corresponding to a minimum detectable fractional absorption (multimode) of 4 x 10(-6). Optimization of the detection method should permit real-time measurement of concentrations below 0.1 ppm. The low cost and convenient operating characteristics of InGaAsP diode lasers make them attractive for many applications in high-resolution near-IR spectroscopy and trace species detection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.