Abstract
Radiation absorbed dose microcalorimetry is a direct technique for energy deposition measurements in low power benchmark mixed neutron and gamma-ray radiation fields. MkII adiabatic microcalorimeters, developed at Imperial College, are described. The results of recent electrical calibration and 60Co gamma-ray field measurements are presented for graphite and iron microcalorimeters. In order to assess sources of error, a finite-difference conduction-heat-transfer computer code has been employed to model the microcalorimeter performance. Model predictions are in good agreement with experimental results. As a result of the error assessment, a redesigned MkV graphite microcalorimeter has been constructed, and results for electrical calibration and 60Co gamma-ray measurements are presented. Systematic errors are reduced by a factor of about 10 for electrical calibration. for the 60Co measurements, the discrepancy between the graphite MkV microcalorimeter and calibrated ionization chambers is less than 2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.