Abstract

We discuss the measurement uncertainty in the presence of different types of fundamental noise processes, especially 1/f noise. The treatment is based on the Allan variance of an n-sample average with two deadtime parameters relevant in metrology. Some results applicable to situations that are common in metrology, e.g., international comparisons, are derived. Experimentally, the output of Zener voltage standards in the frequency range from 10/sup -6/ Hz to 10/sup 3/ Hz is studied and behavior closely resembling that of 1/f noise is found at all time scales of practical interest. The noise floor of a 10 V Zener standard was found to be 60 nV and the 1/f corner frequency 3.5 Hz. Environmental variations cause the long-term (>100 s) deviations of the Zener voltage to slightly increase from the limiting 1/f noise level.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.