Abstract
Accurate clinical sensors and devices are essential to support optimal medical decision-making, and accuracy can be demonstrated through the conduct of clinical validation studies using validated reference sensors and/or devices for comparison. Typically unmeasurable, the true reference value can be substituted with an accepted physiological measurement with an associated uncertainty. We describe a basic model of measurement uncertainty that specifies the factors that may degrade the accuracy of an observed measurement value from a sensor, and we detail validation study design strategies that may be used to quantify and minimize these uncertainties. In addition, we describe a model that extends the observed measurement uncertainty to the resultant clinical decision and the factors that may impact the uncertainty of this decision. Clinical validation studies should be designed to estimate and minimize uncertainty that is unrelated to the sensor accuracy. The contribution of measurement observation uncertainty to clinical decision-making should be minimized but also acknowledged and incorporated into the clinical decision-making process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.